Inhomogeneity
 
Inhomogeneity is the degree of lack of homogeneity, for example the fractional deviation of the local magnetic field from the average value of the field. Inhomogeneities of the static magnetic field, produced by the scanner as well as by object susceptibility, is unavoidable in MRI. The large value of gyromagnetic coefficient causes a significant frequency shift even for few parts per million field inhomogeneity, which in turn causes distortions in both geometry and intensity of the MR images.
Manufacturers try to make the magnetic field as homogeneous as possible, especially at the core of the scanner. Even with an ideal magnet, a little inhomogeneity is always left and is caused in addition by the susceptibility of the imaging object. The geometrical distortion (displacement of the pixel locations) are important e.g., for some cases as stereotactic surgery. Displacements up to 3 to 5 mm have been reported. The second problem is the undesired changes in the intensity or brightness of pixels, which may cause problems in determining different tissues and reduce the maximum achievable image resolution.
mri safety guidance
Image Guidance
General strategies for reducing field inhomogeneity induced artifacts:
Increasing the strength of the gradient magnetic field.
Decreasing the echo time.
Improving the image resolution. Phase encoding. Postprocessing.
spacer
 
• Related Searches:
    • Single Volume Spectroscopy
    • Susceptibility
    • Gradient Echo
    • Susceptibility Artifact
    • Partial Fourier Technique
 
Further Reading:
  News & More:
Why non-magnetic capacitors matter in medical imaging
Wednesday, 19 February 2020   by www.medicaldesignandoutsourcing.com    
Implementation of Dual-Source RF Excitation in 3 T MR-Scanners Allows for Nearly Identical ADC Values Compared to 1.5 T MR Scanners in the Abdomen
Wednesday, 29 February 2012   by www.plosone.org    
MRI Resources 
Knee MRI - Case Studies - Stent - Musculoskeletal and Joint MRI - Manufacturers - Raman Spectroscopy
 
Localized Magnetic Resonance
 
(LMR) A particular technique for obtaining NMR spectra, for example, of phosphorus, from a limited region by creating a sensitive volume with inhomogeneous applied gradient magnetic fields, which may be enhanced with the use of surface coils.
spacer
MRI Resources 
Anatomy - Spectroscopy - Functional MRI - MRI Centers - Online Books - Corporations
 
Pulsed Gradients
 
Briefly applied magnetic field gradients.

See also Bipolar Gradient Pulse and Gradient Magnetic Field.
spacer

• View the DATABASE results for 'Pulsed Gradients' (2).Open this link in a new window

Searchterm 'Gradient Magnetic Field' was also found in the following services: 
spacer
News  (1)  Resources  (3)  
 
Slice SelectionForum -
related threads
 
Exclusive excitation of spins in one slice performed by the coincident combination of a gradient magnetic field and a narrow bandwidth or slice selective RF pulse at a specific Larmor frequency.
spacer

• View the DATABASE results for 'Slice Selection' (8).Open this link in a new window

 
Further Reading:
  Basics:
RARE
Monday, 3 December 2012   by www2.warwick.ac.uk    
  News & More:
Magnetic resonance imaging
   by www.scholarpedia.org    
MRI Resources 
Equipment - MRA - Blood Flow Imaging - Non-English - Quality Advice - Implant and Prosthesis
 
Magnetic Field Gradient
 
Magnetic field gradients are used to change the strength of the magnetic field Bo in a certain direction. Gradients are used in MR imaging with selective excitation to select a region for imaging and also to be able to encode the location of MR signals received from the object being imaged. The field strength is measured in Tesla per meter (T/m).
spacer

• View the DATABASE results for 'Magnetic Field Gradient' (28).Open this link in a new window

 
Further Reading:
  Basics:
Magnetic Field
   by hyperphysics.phy-astr.gsu.edu    
MRI Resources 
Collections - Cochlear Implant - Mobile MRI - Used and Refurbished MRI Equipment - Non-English - Liver Imaging
 
previous      6 - 10 (of 27)     next
Result Pages : [1]  [2]  [3 4 5 6]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 29 April 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]